4) OPERATION
ACA Current clamp-on function
Set the slide-switch function selector to the
clamp jaws for non-invasive ACA current measurements.
CAUTION
●Press the jaw trigger and clamp the jaws around only one single conductor of a circuit
for load current measurement. Make sure the jaws are completely closed, or else it will
introduce measurement errors. Enclosing more than one conductor of a circuit will
result in differential current (like identifying leakage current) measurement.
●Adjacent current-carrying devices such as transformers, motors and conductor wires
will affect measurement accuracy. Keep the jaws away from them as much as possible
to minimize influence.