Dillon EDjunior Benutzerhandbuch - Seite 14
Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Messgeräte Dillon EDjunior herunter. Dillon EDjunior 20 Seiten. Dynamometer
Safe Operation
If you overload this dynamom-
eter you could suffer severe
injuries or death. The total load
on the dynamometer should
NEVER exceed the rated
capacity.
14
Keep all the following in mind as you use the EDjr dynamometer.
The system capacity is equal to the rating of the dynamometers. The
shackle rating should not be used to determine lift capacity of the system.
Any zeroed deadload must be considered as part
of the ultimate load.
Although this instrument has a substantial overload protection rating, the
instrument should not be used above the rated capacity. Doing so can
significantly impact fatigue life of the instrument and cause premature and
abrupt failure. If a higher capacity reading is needed, Dillon insists that a
larger instrument be used.
Safety is always a concern in overhead lifting and tensioning applications.
To limit your liability always insist upon factory supplied shackles and pins
and factory tested and certified safe optional equipment. All DILLON prod-
ucts are designed to meet the published Safe Working Load (SWL) and
Ultimate Safety Factor (USF) standards of the United States Military.
Do not grind, stamp or deform the metal on the dynamometer body in any
way.
Any significant damage or deformation to the loading element is cause for
evaluation by Dillon, particularly in the element side members to the right
and left of the display.
Relieve all torsional and off axis loads.
Apply load in the center of the shackle bow with this instrument.
Off center loading results in substandard performance.
Instrument requires time to stabilize when changing temperatures.
Use only the hardware supplied with this instrument. If no hardware was
supplied, insure that the mating pin and shackle bow is equivalent to the
hardware used at calibration. Otherwise substandard performance can
result.
Dillon recommends only using qualified rigging hardware and cannot
be responsible for unapproved hardware.
This instrument is not designed for the following:
• Applications that see rapid, dramatic temperature swings or thermal
shock. Wide variation in readings can occur.
• Environments with high electromagnetic fields such as cranes employ-
ing electromagnets to lift metal. These induce trace voltages that are
picked up within the load cell lead wiring and appear as inaccurate
loads.
• Intrinsically safe environments. This unit has not been Factory Mutual
tested.
EDjunior Dynamometer User's Manual