Dinel GPLS-25N-0 Gebrauchsanweisung - Seite 5

Blättern Sie online oder laden Sie pdf Gebrauchsanweisung für Schalter Dinel GPLS-25N-0 herunter. Dinel GPLS-25N-0 12 Seiten. Thru-wall level switches
Auch für Dinel GPLS-25N-0: Gebrauchsanweisung (14 seiten)

Dinel GPLS-25N-0 Gebrauchsanweisung
u
NSuiTablE raNgE OF aPPliCaTiON
The sensor also is not intended for level measurement on the gauge-pipe and the
container wall with antistatic treatment (partially electrically conductive).
P
rOCEdurE FOr PuT SENSOr iNTO OPEraTiON
1. i
NSTallaTiON
a) The sensor type GPLS–25N–0, which is intended to point level
detection on plastic or glass and gauge-tubes.
The  sensor  is  fixed  to  the  gauge 
pipe or tube by means of two
plastic straps (2.5 mm width). The
cable should be vertically down-
wards oriented. The maximum wall 
thickness of the tube depends on
the detected medium (see technical
data), the maximum is 8 mm. 
b) The sensor type GPLS–25N–1, which is intended to thru-wall level
sensing of liquids in plastic or glass vessels with flat walls.
The sensor is installed on a clean
and degreased surface of the ve-
ssel wall. The attachment is done
by doublesides adhesive layer.
Orientation of the sensor can be
arbitrary. Maximum thickness of the 
vessel wall depends on the detec-
ted medium (see technical data),
the maximum is 8 mm. 
Plastic straps
4