5. Maximum scale capacity and increment setting
888888
div.deC
CapaC
Examples:
For a 60000kg scale, with 2kg increment:
div.deC
=
2
CapaC
=
60000
*
Increment = the amount that the scale will increment by as weight is added or removed.
1
*
Maximum capacity = the maximum weight that can be measured using the scale you are creating.
2

6. Theoretical calibration

888888
...
Cel.Cap
Cel.sen
dead.ld
div.deC
CapaC
1. Set
and
(see section 4)
dead.ld
4. Enter in
step. The display
shows the theoretical dead load
value. Modify the value and/or
confirm with
.
7. Zeroing of the dead load (pre-tare zeroing)
888888
...
0.Calib
This functionality allows to zero the weigh of the scale structure (e.g.: empty silo, empty vessel, etc.) shifting the scale adjustment points.

8. Zero adjustment

888888
...
zer
1. Unload the scale
Set the decimal point position and the minimum scale increment*
(
0,001
-
0,002
-
0,005
-
0,01
Set the maximum scale capacity*
For a 10000 g scale, with 0,1g increment:
div.deC
=
0, 1
CapaC
=
10000,0
Set the total load cells capacity (up to 999999)
Set the load cells mV/V average (up to 99.99999)
Dead load weight (from -9999.9 to 99999.9)
Cel.Cap
2. Set in
capacity (sum of the nominal load cell
capacities)
5. Save calibration (Press
times until
then press
Zeroing of the deal load (pre-tare zeroing)
Zero adjustment
2. Enter the
zero point
TECH_MAN_ENG_DGT1S_QSG
-
0,02
-
0,05
-
0,1
-
0,2
-
0,5
2
(max
999999
)
the total load cells
C
key many
save?
message will appear,
to confirm)
zer
step to adjust the
1
-
1
-
2
-
5
-
10
-
20
-
50
)
For a 3000kg scale, with 0,05kg increment:
div.deC
=
0 , 05
CapaC
=
3000,00
Cel.sen
3. Set in
the load cell mV/V
average value:
LC1 mV/V + LC2 mV/V + LC3 mV/V + LC4 mV/V
4
3. Save adjustment (Press
save?
times until
message will
appear, then press
to confirm)
C