Cal Test CT3687 Benutzerhandbuch - Seite 10

Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Tester Cal Test CT3687 herunter. Cal Test CT3687 17 Seiten.

instrument and the probe turned on for 30 minutes to
stabilize.
4.
Press [Default Setup] and [Auto] on the oscilloscope.
5.
Press the Channel 1 button, then press the Probe softkey
and set the attenuation to match that of the probe.
6.
Set the oscilloscpe to DC coupled mode and the scope
offset to 0 volts.
7.
Set the oscilloscope to average mode (x16) or high-
resolution mode to reduce oscilloscope noise.
8.
Using the offset adjustment tool (included), adjust the probe
offset voltage to 0 volts
4 Cleaning
This probe does not require any particular cleaning. If necessary, clean
the case with a cloth slightly moistened with soapy water. Make sure
the probe is completely dry before reconnecting it to an oscilloscope.
WARNING
Dry the probe thoroughly before attempting to make voltage
measurements.
CAUTION
Do not subject the probe to solvents or solvent fumes as these can
cause deterioration of the probe body and cables.
800-572-1028
caltestelectronics.com
9