Error Messages

Error Message
Sn or PL
rP
H
Lo

Understanding The Results

The result that displays on this device is an estimate of the subject's
BAC. The Department of Transportation (DOT) employs the standard
of 0.02% as a threshold for a positive test for alcohol in no tolerance
screening. Nearly all states have 0.08% as the legal BAC limit for
tasks such as operating a motor vehicle. Some states prohibit driving
commercial vehicles or any vessel with a 0.04% BAC or higher. This
device should only be used as a screening device and may only give
a reading of the possible presence of alcohol in the blood of the test
subject. Correlation between breath alcohol content and BAC depends
on many factors, including temperature and health conditions. The
exact BAC in the blood of the test subject cannot be exactly
determined by using a breath alcohol screening device. Never
drink and drive.
Cause and Course Of Action
There has been an error with the sensor. This can occur if
the test subject does not wait at least 20 minutes after
smoking, eating or drinking before using the unit. or if
the unit has not been used for an extended period of time.
Press the Power On/Off button once to turn off the unit
and repeat Steps 1 – 6 above. If this message is
displayed four consecutive times, the unit may need
serviced.
Repeat the test because the unit did not detect an even
alcohol concentration in the breath sample. This can
occur if the user varies the blowing force when performing
a test or if the exhaust port is obstructed. Repeat Steps
1 – 6 above to perform another test.
The unit detected a very high alcohol concentration
(above 0.40% BAC).
Low Battery. Replace AA batteries.