Alcohol Countermeasure Systems ALCOSCAN Handbuch Kalibrierung - Seite 12

Blättern Sie online oder laden Sie pdf Handbuch Kalibrierung für Prüfgeräte Alcohol Countermeasure Systems ALCOSCAN herunter. Alcohol Countermeasure Systems ALCOSCAN 16 Seiten. Passive breath alcohol tester

ENTERING ADMIN MODE

The ALCOSCAN tester must be in ADMIN Mode in order to calibrate. Follow these
steps to enter ADMIN Mode:
1.
Press the power button 4 times in quick succession and hold (click, click, click,
click and hold). Release the power button when 1 beep is heard.
A combination of LED lights will illuminate. Each light indicates a number, as
shown in the diagram below.
2.
Add up the numbers indicated by the illuminated LEDs. Take that sum and press
the power button that many times in quick succession.
The WARN LED light will blink blue when the appropriate number is entered. The
ALCOSCAN tester is now in ADMIN Mode and can be calibrated.
Three beeps indicate that the number entered is not correct.
Continue to the next section.
7
ALCOSCAN