Troubleshooting
WARNING DISPLAY "Flow": If you see "Flow" in the display, the breath sample was not strong or
long enough for measurement. Please try again and make sure the breath is strong and steady
during sampling. As soon as you start blowing into the device, you will hear a long beep indicating
how long you should blow.
WARNING DISPLAY "bAt": If you see "bAt" in the display, the battery power is low. Please replace
the batteries with a new pair. Make sure to properly align the batteries by looking at the +/- marks.
WARNING DISPLAY "Err2/Err4": If either of these codes appear in the display, the device is
reporting a general problem with the alcohol sensor, or that the device has been physically
damaged. Please replace sensors or contact AK GlobalTech Corp.
WARNING DISPLAY "Er
improperly connected or missing altogether. It is very important when installing a
sensor module that the 9-pin contacts are securely in position and that the central tab
is locked in place.
WARNING DISPLAY "Err9": Power source is too weak to complete operation. Please replace
batteries with a new pair.
For the other troubles, please contact AK GlobalTech at [email protected], along with your
error codes and photos if available.
Alcohol Sensor Advanced Electro-Chemical Fuel-Cell
Detection Range 0.000 -0.400 g/dL (Accuracy to 0.02)
Warm-up Time 5 seconds - 2 minutes
Response Time Within 10 seconds at 0.100%BAC
Calibration In lieu of traditional calibration, the sensor module must be calibrated every
Breath Sampling Sample complete on audible click
Sensor Accuracy +/-0.005 up to 0.100 g/dL
Display 4-Digit LCD
Battery Two 1.5V AAA Alkaline (included)
Weight 3.0 oz
Dimensions 4.0 x 2.0 x 0.7 in
Casing Gray ABS Plastic
Operating Temperature 32-104 degrees Fahrenheit or
Warranty One (1) Year (Refer to "Warranty" section
r6": The device is reporting that the sensor module is either
- Specifications -
1,000 tests or at least once yearly, whichever comes first, in situations
directly overseen or mandated by the DOT, users must perform calibration
checks according to the manufacturer's QAP for the Revo TS200.
0-40 degrees Celcius