After pressing the power
switch a third time, the display
is horizontal and upright
After pressing the power
switch a fourth time, the
display is horizontal and
inverted
After pressing the power
switch a fifth time, the display
is vertical and facing away from
the resident
After pressing the power
switch a sixth time, the display
is vertical and facing toward
the resident
Calibration & Accuracy Verification
• The device cannot be calibrated in the field by the
end user.
• If you suspect the device is providing inaccurate
results, use a second device to verify accuracy. Do
not rely on results you suspect to be inaccurate.
• Immediately discontinue use of the device if it
provides inaccurate readings.
Maintenance, Storage & Disposal
• Replace the batteries as soon as the low-battery indicator
on the display screen is lit.
• Clean the surface of the device before it is used in
diagnosis of residents.
• Remove the batteries if the device will not be used for an
extended period of time.

Limited Warranty

We offer to you, as the original purchaser, a warranty for the Attendant Fingertip Pulse Oximeter. Our warranty
applies for the limited warranty period stated below. If any device or device part listed below is defective in
material or workmanship during the applicable limited warranty period, we will repair or replace it at our cost.
Please note that the decision to repair or replace a device or device part will be at our discretion. Our warranty
applies only if the device is properly maintained by the original purchaser for normal, indoor use and does not
cover normal wear and tear, modification of the device, or damage caused by abuse, improper use, failure to
maintain, use which exceeds the published device limitations, or the combination of any device with another
product. In addition, our warranty does not cover fading, colorfastness, stains, spills or exposure to chemicals,
odors, heat or light. In certain cases, we may provide you repair or adjustment instructions and/or replacement
parts and ask you to perform a repair or adjustment or replace a defective part.
Our warranty gives you specific legal rights, and you may also have other rights, which vary from state to state.
Please note that our limited warranty period begins when we ship the device to you. The limited warranty period
and our obligations under the warranty end once you transfer the device to someone else or at the end of the
applicable limited warranty period identified below, whichever is earlier.
Attendant Fingertip Pulse Oximeter
(excluding batteries and accessories)
• Store the device in a place where ambient temperatures
range from 14° to 104°F (-10° to 40°C). A low-humidity
environment is recommended.
Cleaning
• Use a soft cloth dampened with 70% isopropyl alcohol
to clean the rubber that comes into contact with a finger
during testing. Also clean the test finger using alcohol
before and after each test.
• CAUTION – Do not pour or spray any liquids
onto the device, and do not allow any liquid to enter
any openings in the device. Allow the device to dry
thoroughly before reusing.
Follow local ordinances and recycling instructions regarding
disposal or recycling of the device and device components,
including batteries.
Measurement Principle
The Oximeter is calibrated to display functional oxygen
saturation using the principles of the Lambert Beer Law,
according to Spectrum Absorption Characteristics of
Reductive Hemoglobin(R Hb) and Oxyhemoglobin (O
Hb) in red and near-infrared zones. Using Photoelectric
Oxyhemoglobin Inspection Technology and Capacity
Pulse Scanning and Recording Technology, the Oximeter
measures SpO
and pulse rate by shining two beams
2
of differing wavelengths of light (660nm red and 940nm
near infrared) through the tip of the fingernail. The
Oximeter measures the amount of each wavelength
of light that passes through the fingernail using a
photosensitive element. The microprocessor analyzes
these measurements, determines the SpO
pulse rate of the resident, and displays this information
on the LCD display.
Warranty Period
(Parts)
2 years
1-800-634-7328
level and
2
Anticipated Usable
Device Life
2 years