Nikon Labophot Gebrauchsanweisung - Seite 14

Blättern Sie online oder laden Sie pdf Gebrauchsanweisung für Mikroskop Nikon Labophot herunter. Nikon Labophot 28 Seiten. Y-r stand biological microscope

Nikon Labophot Gebrauchsanweisung
3.
Condensers
1) Abbe condenser
NA
= 1.25. This is used with 4X -
100X ob-
jectives. The graduation
of
this
condenser
indicates the diameters in mm of the aperture
diaphragm opening.
2) Swing-out Achromat condenser
N.A.
=
0.9. Dry system.
It is used in combination with objectives from
2X to 100X, and provided with a swing-out top
lens which is to be swung out when using the
2X or 4X objective. Its adjustable aperture scale
is graduated in N.A. ratings.
3) Achromat/aplanat condenser
NA
=
1.35. Oil system.
The spherical, coma and ch rom<ltic aberrations
being ideally
corrected,
this large aperture
condenser is used with 20 X - 100 X ob jectives.
The standard thickness of glass slide should be
1.2mm.
Apply oil between the condenser and glassslide.
It
is recommended
that
this condenser be
employed especially in combination
with the
Plan Apochromat
objectives. When using the
lOOX objective for observation in combination
OPTICAL PATH
Collector
lens
Halogen
lamp
14
with the CFW lOX eyepiece, it is possible to
close the field diaphragm down to 45% of the
viewfield.
4) Darkfield condenser (Oil)
N.A.
=
1.43 -
1.20. Oil system. Used in dark-
field
microscopy.
Apply
oil
between
the
condenser and glass slide. (It is recommended
to use a thinner glass slide.)
This condenser is used in combination with the
objectives
lOX
-
lOOX with aperture dia-
phragm (N.A.:
up to 1.11.
5) Darkfield condenser (Dry)
N.A.
=
0.95 - 0.8. Dry system. Used in dark-
field microscopy. Magnifying powers of usable
objectives are lOX - 40X (N.A.:
up to 0.7).
4.
Illumination System
(Fig. 15)
The optical
system for
illumination
in the
LABOPHOT microscope is constructed to fulfill
the Koehler illumination
requirements perfect-
ly, and offers a bright, uniform
field without
any change-over manipulation.
As a standard light source, use the Halogen
lamp 6V 20W (PHI LIPS 7388).
Objective
Filter
Field lens
Fig. 15