Armstrong Medical AquaVENT FD140 Leitfaden für die Verwendung

Blättern Sie online oder laden Sie pdf Leitfaden für die Verwendung für Medizinische Ausrüstung Armstrong Medical AquaVENT FD140 herunter. Armstrong Medical AquaVENT FD140 2 Seiten. High flow oxygen therapy

Guidance for Use
1. Select correct size of AquaNASE®
nasal cannulae.
2. Apply AquaNASE® nasal cannulae and
secure neck strap as shown overleaf.
3. Insert humidification chamber onto heater
plate and spike water bag.
4. Attach circuit as shown a-d.
2
*This guidance should be read in conjunction with the IFU supplied with the specific Breathing Circuit, Heater Humidifier, AquaNASE® nasal cannulae and the Flow Driver.
*
1
4d
High Flow Oxygen
Therapy – FD140
5. Connect heater cable and
temperature probes.
6. Turn on heater and select invasive mode.
7. Set flow at 20L/min and select prescribed
FiO
for 2 minutes.
2
8. Increase flow to max 70L/min as required
monitoring SpO
FD140 Flow Driver
(Supplied
Separately)
5
Heater Humidifier
(Supplied
Separately)
and respiratory rate.
2
AquaVENT®
FD140
0
5 10 15 20 25
PAP
0
2
8
7
4a
4b
4c
3
6
PMA_AMFD140_HFOT_DSIFU_v1 UK 01/12/17_15639