CAUTIONS
Annual Servicing
Daily Testing
Performance
IMPORTANT
Loss of Signal
Tip Switch
DFU-0320-3 Rev. 0
For continued safety and performance of the switch, it
is recommended that the functionality and electrical
safety of the switch be verified on an annual basis by an
authorized representative.
It is essential that the footswitch be inspected every day
or before use.
The RF wireless footswitch system operates within the
2.4GHz ISM Radio Frequency Band. It should be tested
to assure compatibility with any device it is connected
to or environment it is working in.
Report all problems experienced with the footswitch. If
the footswitch is not working properly, contact your
service representative for service. The footswitch
should not be used if it is not working properly.
If the receiver does not receive a signal from the
transmitter for a period of 500ms it will shut off the
outputs. However, once the transmission is restored the
receiver will function normally.
The RF wireless foot switch operates in a 360 degrees'
mode of operation. However, every application is
different. You may need to adjust the position of the
receiver or transmitter to obtain optimal performance.
Up to 5 transmitter and receiver pairs may be operated
in a single room.
If the footswitch is tipped 60 degrees or more or
disturbed in such a way as to activate the internal tip
switch, the receiver's output will shut off, and the
footswitch will be inoperable until it is returned back to
the normal orientation and stabilized.
8 of 20