11. Checklist of safety and function test
ARTROMOT®- K4
Safety test
Protective earth conductor resistance
Ground leakkage current EN 60601 / IEC 601/ VDE 0751
Or
Ground leakkage current as in UL 2601
Function test
1. Switch on the ARTROMOT® K4. Press the two outer buttons simultaneously.
Display: Software version VX.X XX.XX.XX (X = optional) Keep on pressing.
Display: ARTROMOT K4
2. The maximum range of motion for Extension/ Flexion is -10 to 125 degrees.
Check the angle in position 0 degrees. Tolerance +/- 5 degrees.
Check the angle in position 60 degrees. Tolerance +/- 5 degrees.
Check the angle in position 100 degrees. Tolerance +/- 5 degrees.
3. Check the emergency-off function.
Start the ARTROMOT® K4 in any mode.
Press any key, the ARTROMOT® K4 will stop immediately.
Check this for all keys.
4. Set up the special function "new patient"
Press START. The ARTROMOT® -K4 will move automatically to 30 degrees
and stop there.
5. Check the set values of "new patient"
Extension
Flexion
Speed
Reverse
Therapy Duration
6. Start the ARTROMOT® -K4 in the motion range between –10 to 120 degrees.
Set the speed
Both extreme points should be reached within 70 – 100 seconds.
7. All special funtion are deactivated. Select a therapy time of 3 minutes.
At the end of the 3 minutes the ARTROMOT®-K4 will stop automatically
in the set up value of extension + 10 degrees.
"Product version"
->
Display: 25°
->
Display: 35°
->
Display: 100 %
->
Display: 25
->
Display: 00:00
to 100%.
Measured value
0,1 Ohm
500 µA
300 µA
press to activate it.
17
Date/ Signature
Ohm
µA
µA
OK