9.- Technical specifications

AKO-55624 / AKO-55724
Power supply ......................................................................................................................100-240 V~ 50/60 Hz
Maximum input power..................................................................................................................................15 W
No. of inputs
AKO-55624 ......................................................................................................................1
AKO-55724 ......................................................................................................................4
Compatibility of inputs .....................................................................................................AKO-575xxx transmitter
..............................................................................................................................................AKO-5761 detector
Relays.....................................................................................................................................SP T D 8(2)A 250 V~
Working ambient temperature .........................................................................................................-5 ºC to 50 ºC
Storage ambient temperature ........................................................................................................-30 ºC to 60 ºC
Degree of protection.....................................................................................................................................IP 65
Installation category ....................................................................................................................II s/ EN 61010-1
Degree of pollution......................................................................................................................II s/ EN 61010-1
Double isolation between power supply, secondary circuit and relay output.
Sound power..........................................................................................................................90 dB(A) at 1 metre
MODBUS address..........................................................................................................................Shown on label
Dimensions...........................................................................................290 mm (W) x 141 mm (H) x 84.4 mm (D)

10.- Accessories

Calibration tool for AKO-58110 transmitters
Allows for verification (bump test), field calibration and reset to zero of AKO-575xxx gas transmitters.
14
5572H402 Ed.03