the hose from the Pump to speed up the air release.
the hose from the Pump to speed up the air release.
the hose from the Pump to speed up the air release.
the hose from the Pump to speed up the air release.
from the Pump to speed up the air release.
..
..
..
..
5.2.3 Audiovisual Alarm
l Power Failure – When an electrical shortage occurs or power cord is unpluged
without turning off the Pump or
unintentionally), the "Power Failure" indicator will light up along with buzzer
and will last 20 minutes.
NOTE: When the Pump has not been used for more than 3 months, the Pump
may need 6 hours or more of operating time for the alarms to function properly.
l Low Pressure –When an abnormal low pressure occurs in body section for 2.5
minutes after the pressure is below the abnormal pressure threshold, the
"Low Pressure" indicator will flash and beep.
and fault conditions continue, the alarm will resume.
l Service (Malfunction) –When fault conditions occur, the "Service" indicator
will light up along with buzzer.
Note: Refer to Table 2 for Warning Code Reference Table if error code
appears on the display or refer to 10. TROUBLESHOOTING.
5.2.4 Alarm Mute
l When alarms are triggered, both LED light and buzzer will turn on to warn the
patient and caregiver. By pressing the button, it will temporarily mute the
buzzer so the caregiver may check for possible causes. Should the situation
not resolved and fault conditions continue, the alarm will resume. Refer to 10.
TROUBLESHOOTING for diagnosis.
l During "power failure", pressing "alarm mute" will cease all buzzers and
indicators and turn off the system.
l During "low pressure alarm" if the pressure resumes back to normal then the
low pressure alarm will stop.
l When more than one alarm is triggered, the alarm will be performed
according to priority level. Refer to Table 2 Warning Code Reference Table for
priority level.
is pressed (intentionally or
Should the situation not resolved
10
11
11
11
11