Etac Molift Partner 255 Benutzerhandbuch - Seite 14
Blättern Sie online oder laden Sie pdf Benutzerhandbuch für Mobilitätshilfe Etac Molift Partner 255 herunter. Etac Molift Partner 255 15 Seiten.
Molift Partner 255 / www.etac.com
Cleaning and disinfection
Clean on a regular basis. Clean surfaces with
a damp cloth using an appropriate pH-neutral
detergent. Do not use solvents or strong liquids,
this may damage surfaces on the lifter. For
disinfection when needed; use isopropyl alcohol.
Avoid abrasive cleaning products. Check
emergency stop and emergency lowering after
cleaning. The lift should not be exposed to run-
ning water.
Make sure not to damage or
remove labels when cleaning.
Periodic Inspection Scope
Periodic inspection is a visual examination (par-
ticularly of the hoist's load bearing structure and
lifting mechanism with attachments, brakes, con-
trols, safety devices and person-support devices)
according to Periodic Inspection Report for Molift
Air. This can be found on www.etac.com.
Periodic Inspection shall be performed at least
once a year or more frequently if required by local
requirements. The inspection must be performed
by service personnel authorized by Etac. Contact
Etac for training and authorization or recommen-
dation of an approved service partner.
When performing a periodic inspection, the
inspector shall fill out the inspection report for
Molift Air. The reports should be retained by the
person(s) responsible for servicing the hoist. If
the inspection reveals defects and damages, the
owner shall be notified and a copy of the report
should be sent to Etac, [email protected].
In the event of danger to safety,
the hoist shall immediately be
taken out of service and marked
clearly with "out of order" and
shall not be used until the hoist is
repaired
Maintenance
Reconditioning
Follow cleaning and installation procedure, com-
plete periodic inspection and use checklist after
installation to recondition the lifter.
Recycling
Refer to "Recycling instructions" for how to
properly dispose of product. This can be found on
www.etac.com.
Spare Parts
A list of spare parts is available on request.
Periodic Inspection
When periodic inspection is completed the
inspector shall mark the hoist with a sticker on
the control label showing the date when periodic
inspection is performed. The control label can be
found on the inside of the Side cover, and this
will then indicate when next service should be
performed.
CONTROL
12 Month
13
Mark label
with month
and year of
inspection