rOTOr cOnFigurATiOns:

This rotor is designed to hold up to 6 tube carriers.
Follow the procedure below for proper tube loading.
Six (6) Carrier Horizontal Rotor (standard):
Your centrifuge must contain a balanced load in order
to work properly. to ensure that the load is balanced,
keep these rules in mind when inserting test tube
samples into the four (6) position horizontal rotor:
1. Opposing tube carriers must be identical.
2. Opposing tube carriers must be empty or loaded with an equal number of
equally weighted samples.
3. If an odd number of samples is to be spun, use a water-filled tube to mate
with the unpaired one.
1. Opposing tube carriers must be identical.
2. Opposing tube carriers must be empty or loaded with an equal number of
equally weighted samples.
3. If an odd number of samples is to be spun,
use a water-filled tube to mate with
the unpaired one.
4. (Four Position Carriers Only):
When loading tubes into opposing
carriers, the carriers will be
balanced when the tubes are loaded
diagonally from one another
(not across from one another).
For example: six tubes are to be loaded
into two carriers. In each carrier, two tubes
are placed in the bottom and one tube is
placed in the upper left-hand position.
See picture at right.
note: Do not load more than two (2) tubes per carrier if the tubes are greater than
100mm long (15mL). When loading these tubes into the carriers, place them all
on either side. For example, to load twelve (12) 15mL tubes, place two tubes on
the right-hand side of each carrier.
WARninG:
When using glass test tubes, place a 7mm cushion at the bottom of each
location in the carriers. Failure to do so may result in damage to the test tube.
Six (6) Place 50mL Fixed-angle Rotor (optional):
This rotor is designed to hold up to six (6) 50mL tube holders.
Your centrifuge must contain a balanced load in order to
work properly. to ensure that the load is balanced, keep
these rules in mind when inserting test tube samples into
the 6 place, 50mL fixed-angle rotor:
1. Opposing tube holders must be identical.
2. Opposing tube holders must be empty or loaded equally weighted samples.
3. If an odd number of samples is to be spun, use a water-filled tube to mate
with the unpaired one.