UNIT OPERATION:
TA-90D Air/Water Syringe
and the left button for water operation. Depressing both buttons will create a
mist. The syringe features quick-change autoclavable tips: To remove a tip,
press on the locking collar surrounding the tip socket and pull the used tip
straight out of the socket. To insert a new tip, press locking collar and push tip
into socket as far as it will go. Release collar and gently tug on tip before
using to ensure that tip is securely locked into socket.
Syringe Tip Sterilization:
1) Remove contaminated syringe tip.
2) Remove all visible signs of contamination
before autoclaving.
3) Autoclave tip at 132° C (270° F) for ten
minutes.
4) Sterilize between each patient use.
NOTE: Since only the tips can be autoclaved, it is
recommended that the air/water syringe be bagged
with a disposable, single-use plastic sleeve
between each patient use.
AA-21SX Air Vacuum System
compressed air source, and the vacuum bottle assembly is attached to the
air receptacle on the side of the unit case, the Air Vacuum System should be
fully functional. Insert a standard disposable vacuum tip into the hose end.
Depress button on hose end to activate vacuum. Observe waste bottle while
in use and dispose of waste when bottle is approximately 3/4 full.
ON/OFF
BUTTON
Avoid waste flow-back during use by keeping the ON/OFF button depressed
prior to entering the oral cavity and until the hose end has been removed
from the patients mouth. Clean the entire hose assembly and waste bottle
after each use by evacuating a cleaner/disinfectant through the system and
thoroughly scrubbing entire assembly. CAUTION: Use only NON-foaming
cleansers in the vaccum lines.
Handpiece Control-
piece control with water coolant and pressure gauge. Attach any 4-hole
Midwest style highspeed, lowspeed, or air sonic scaler to the control tubing
and place in the holder on the front of the unit. Depress the disc foot control
for handpiece operation. For water coolant, switch the 'Water Coolant' toggle
position
to the
'ON'
HighSpeed
LowSpeed
Air Scaler
-
Depress the right button for air operation,
The ADU-08SXE unit includes one manual hand-
. Water coolant will activate with handpiece operation.
Attach
Handpiece
SYRINGE TIP REMOVAL
SYRINGE TIP REMOVAL
-
When the unit is connected to a
INSERT DISPOSABLE
Depress
Foot Control
DEPRESS
DEPRESS
LOCKING COLLAR
LOCKING COLLAR
VACUUM TIP