V. Troubleshooting Guide
The following general guidelines apply to most PCR reactions. However, no
attempt has been made to address troubleshooting for all of the many applica-
tions for which the Advantage-HF Kit can be used. When using the kit with
another CLONTECH product, additional, application-specific troubleshooting
information can be found in the relevant User Manual.
A. No product observed
PCR component
missing or degraded perform a positive control to ensure that each com-
Too few cycles
Annealing temp.
too high
Suboptimal primer
design
Not enough
template
Poor template
quality
Denaturation temp. Optimize denaturation temperature by decreasing or
too high or low
Denaturation time
too long or too short ing it in 10-sec increments. (A denaturation time that
Extension time too
short
page
Protocol # PT3139-1
14
Version # PR76834
Use a checklist when assembling reactions. Always
ponent is functional. If the positive control does not
work, repeat the positive control only. If the positive
control still does not work, repeat again replacing
individual components to identify the faulty reagent.
Increase the number of cycles (3–5 additional cycles
at a time).
Decrease the annealing temperature in increments
of 2–4°C.
Redesign your primer(s) after confirming the accu-
racy of the sequence information. If the original
primer(s) was less than 22 nt long, try using a longer
primer. If the original primer(s) had a GC content of
less than 45%, try to design a primer with a GC
content of 45–60%.
Repeat PCR using a higher concentration of DNA
(after trying more cycles).
Check template integrity by electrophoresis on a
standard TBE-agarose gel. If necessary, repurify
your template using methods that minimize shearing
and nicking.
increasing it in 1°C increments. (A denaturation
temperature that is too high can lead to degradation
of the template, especially for long target sequences.)
Optimize denaturation time by decreasing or increas-
is too long can lead to degradation of the template,
especially for long target sequences.)
(Especially with longer templates) Increase the
extension time in 1-min increments.
Technical Support TEL:415-424-8222 or 800-662-CLON
FAX: 415-424-1064 or 800-424-1350