Connect a milliamp-meter across 4-20mA OUTPUT1terminals with the correct polarity.
The display will show the corresponding ADC count corresponding to 4mA Output current.
Use theSHIFT key to shift the cursor to desired digit and INC key to change the desired digit from 0 to
9to get 4mA on the milliamp meter.
Press MODE key to save theZERO calibration point of 4-20mA DC for Output 1.
In the same sequential process, the following parameters for each Stream can be set :
Output Zero Adjust for Streams 2 to 6.
High Alarm for Streams 2 to 6.
This completesthe 4-20mA retransmission analog output calibration of the instrument.
Press MODE key to exit from Retransmission Calibration Mode. After pressing the MODE key, the
OLED Display will return to normalPROCESS SCREEN.
SETTING MODE
The ASHE Stream Selector Model STL-96 provides both, AUTO and MANUAL modes of operation. One
can select AUTOor MANUAL mode directly from the front panel key. There are two sub-mode options
in the AUTO mode of operation, viz. Timer and Pulse modes.
Enter the number 06 into the Select Menu by using INC key in the PARAMETER SETTING MODE
SCREEN.
Press MODE key to enter into the Setting Mode.
The OLED Display shows:
Press the SHIFT key to move the cursor to the last digit.Enter Password 0004 by using INC key and
press MODE key. After pressing the MODE key, the OLED Display shows
Use the INC key to select the desired mode of operation to either Timeror Pulse.
Press the MODE key to proceed further. After pressing MODE key, the microcontroller will save
selected mode of operation and move to Sample Time setting.
After pressing the MODE key, the OLED Display shows:
Password
0000
Select Mode
Timer
Start Channel
01
ASHE CONTROLS PRIVATE LIMITED.
S
S
STL-96
TREAM
ELECTOR
Operation Manual