ASI 1045 Betriebs- und Wartungshandbuch - Seite 9
Blättern Sie online oder laden Sie pdf Betriebs- und Wartungshandbuch für Zahnärztliche Ausrüstung ASI 1045 herunter. ASI 1045 20 Seiten. Dental operative unit classic assistant’s cart
c
laSSic
1.9
Optional Suction Connections and Piping Recommendations
ASI supplies suction line tubing made from a special grade of material that is inert to common
caustic irrigants and disinfectants used in dentistry. The suction tubing has a 5/8" inside
diameter (I.D.). Included with suction packages are PVC adapters to connect the 5/8" tubing to
either 1/2" or 3/4" PVC pipe (see Figure 1.9). If using 1/2" copper pipe for suction, the tubing
can simply be slipped over the pipe and clamped into place.
: PVC pipe is recommended for suction plumbing lines and NOT copper,
RECOMMENDATION
as the copper can corrode from some of the caustic irrigating and disinfectant solutions used
in dentistry, such as sodium hypochlorite. Check your local building codes to see which piping
materials may be used for suction lines.
1.10
Pipe Junctions -Housings and Concealment
It is recommended to cover the area where the pipe stubs and
connections are made to prevent them from being bumped into and
to conceal them within the operatory. If a computer or monitor will be
attached to the delivery system, it is advisable to have network cables
run through the same area as the plumbing to manage all the various
tubing, wires and cables. An optional junction box housing is available
from ASI (see figures 1.10A an 1.10B).
This allows the plumbing and electrical to run within the wall between
two studs. A decorative trim panel is then placed over the wall with
cut-outs for the umbilical lines. The location in the wall should be
approximately 12" to 18" off the floor and should be near where the cart will be placed.
Rev.A
ASI - The leader in Advanced Systems Integration www.asimedical.net
a
1044/1045/2044/2045
dvanced
Figure 1.9
Figure 1.10 A
Figure 1.10 B
m
anual
9