The temperature signal is sent to the pressure display using a J-type thermocouple set in a
maleLuer-Lok fitting. The 5-foot length of signal wire ends in a standard 2-pin plug that plugs into the
back of the display unit. The thermocouple is calibrated by the manufacturer for digital displays.
The LED temperature display indicates temperature in °C.
MidJet pressure display unit Instructions for use 18117465 AD
WARNING
If you autoclave or use hot caustic cleaning solution (0.5 N NaOH) to
clean the pressure transducers, the transducers can fail causing
leaks of potentially hazardous cleaning or process fluid and bodily
harm. To avoid leaks of potentially hazardous solutions, do not
autoclave the pressure transducers. Limit exposure to caustic solu-
tion (0.5 N NaOH) to room temperature and only for short periods
(30 to 60 minutes) of time. Rinse all caustic solution thoroughly
from the transducers after exposure.