Flocca Flow Manual de mantenimiento rápido - Página 2
Navegue en línea o descargue pdf Manual de mantenimiento rápido para Instrumentos de medida Flocca Flow. Flocca Flow 2 páginas. The ultrasonic meter measures the depth of flow entering the basin forebay through a defined size pipe or channel.
6. CHECK SETTINGS
»
From Main Screen, go to Settings
»
Check input settings are correct as per setup
7. CALIBRATE ULTRASONIC
»
The ultrasonic is very sensitive and this calibration should be
performed regularly. The value can become inaccurate from the
smallest movement in the pipe or the sensor itself
»
From Main Screen, go to Manual Calibrate
»
Select Calibrate Ultrasonic
»
Enter Actual Depth (mm) as 0
»
Select Update Calibration
»
Actual depth would only ever have a value if there is water standing
in the pipe at time of installation
8. MANUAL (TEST) DOSE
»
Place measuring jug (min. 1 L at outlet hose)
»
From Main Screen, go to Manual/Calibrate
»
Dose Fixed Volume (L) – enter 0.5 L
»
Press Start (No need to press Stop, dosing will stop automatically)
»
This first dose will fill any voids in the outlet assembly and hose
»
Return chemical to IBC and place measuring jug at outlet again
»
Dose Fixed Volume (L) – enter 0.5 L. Press Start
»
No need to press Stop, dosing will stop automatically
»
Check that volume in jar matches fixed volume, within 10%
»
Press Home - finished, your Flocca is good to go
9. PRIOR TO STORAGE
Whilst unit is still setup, it is recommended to rinse out solenoid and
flow meter to remove chemical residue:
»
Close IBC valve and detach camlock
»
Whilst flow meter and solenoid still attached to controller (but not
to IBC), perform dummy manual test dose to open solenoid, using
value of 1L or more
»
With jug pour water through flow meter and solenoid to rinse out
and remove residue
»
Repeat if needed
»
Dry with towel and pack away
flocca.com.au
Flocca Flow - Quick Maintenance Guide
2