Device Specifications
Measuring Range: 
Accuracy (all ranges) 
Power Supply: 
Sensor Ratings: 
Pump Characteristics: 
Outputs: 
 
 
 
1% CO2 0‐10,000 ppm (0‐1% vol.) 
5%, 30%, 65%, 100% CO2 0‐5%, 0‐30%, 0‐65%, or 100% 
Maximum Voltage 30VDC 
Minimum Voltage 18VDC 
Power Consumption ~1Watt avg. (pump running) 
Life Expectancy >15 years 
Maintenance Interval No maintenance required 
Warm‐up Time <1 min (instant 
Maximum Flow (STP) 800 ml/min 
Maximum Vacuum 52% 
Maximum Pressure 600 mbar restart, 150 mbar 
Maximum System Pressure ~1 ATM 
Range 0‐100% vol. 
Output Value 4‐20mA, linearly scaled 
Relay COM, NC, NO 1A @ 30VDC 
RS‐485 Modbus interface 
 
vol., respectively 
±0.5%, ±3% measured value 
 
 
measurements) 
 
continuous 
 
CO2Meter
 