4. OPERATING INSTRUCTIONS
MEASUREMENT
Preparation
1) Make sure that the temperature of the water sample to
be analyzed is between 15 and 35°C to ensure the best
accuracy in the chlorine measurement;
2) Wash the cuvette 3 to 4 times using the water sample
that will be analyzed to set the container, avoiding
reading deviations;
3) Fill the cuvette with 10mL of the sample. Notice the line
in the cuvette body;
4) Close the cuvette with its lid;
5) Clean and dry the outer wall of the cuvette, using a wipe
or soft towel paper, to eliminate residues, grease and
moisture that may interfere with the measurement;
Zero point adjust
6) Press the Power-Measurement button to turn on the
instrument. Will appear Insert reference;
7) Insert the cuvette into the compartment and close the
measurement cover;
8) With the instrument in the VERTICAL position, briefly
press the button. The sample will adjust at zero point;
9) The message: Analyzing reference will appear on
display;
10) Wait until the display shows the message Insert
reagent;