1.4 Major Technical Performance

1. Nominal drift diameter:(4~200)mm, refer to Table 1 for the basic parameters;
2. Medium temperature:(-20~80)°C; Split type (-20~120) °C;
3. Ambient temperature:(-20~70)°C;
4. Precision: ±0.5%, ±1%;
5. Detector signal transmission wiring system: three-wire voltage impulse (three-core
shield wire);
6. Power supply:
1) TB3WE Three Wire
External Power:12~24VDC/30mA(-20%~+15%),if no output can be as low as 6V
Battery:Lithium battery 3.6V(2/13Ah)/0.4mA
2)TB2WE Two Wire
12~24VDC/4-20mA(-20%~+15%)
7. Transmission distance: the distance between the sensor and the display instrument
may be as far as 1000m;
8. Local display power supply: 3.6V (Lithium battery, may be used continuously for
more than 3 years);
9. Display mode: local LCD displays instant flow and cumulative flow;
10. Output Signal
a)TB3WE Three Wire
1)Pulse Output:High level voltage amplitude≥5V, low level<0.5V
2)Three sire 4-20mA linearity correction current output(need ground wire)(load
resistance≤800Ωat 24V)
3)RS485 communication:flowmeter with RS485 interface, communication distance
1200mm.
b)TB2WE Two Wire
1) Two wire 4~20mA linearity correction current output(need ground wire)(load
resistance≤600Ωat 24V)
AKTEK Endüstriyel Ekipman ve Enstrümantasyon LTD.ŞTİ.
İMES Sanayi Sitesi B-Blok 204.Sok. No:40 Y.Dudullu-Ümraniye/İSTANBUL
3