3. Operation
3.5 Sampling/Measuring Mode
IMPORTANT NOTE: Always ensure that the filters provided are used and
replaced regularly
IMPORTANT NOTE: Before each day of use, we recommend that the unit
is zeroed in air by following the "Calibrate in Air" procedure in
Section 4. Calibration.
To turn on the integral gas pump, briefly press the
To turn off the pump, again briefly press the
3.6 Resolution adjustment
To adjust the measurement resolution from 1 decimal place to 2 decimal places tap the
"Percent CO2" area of the screen:
3.7 Measurement procedure
When sampling the CO2 level of a CO2 Incubator, connect the unit tubing to the
sample port and press the
A sample of the atmosphere from the CO2 Incubator will then be drawn in through the
tube and the "PERCENT CO2" reading will start to change as the gas to be measured is
drawn into the unit.
Monitor the reading until it stabilises. This will typically take around 90 seconds.
Take note of the reading and press the
IMPORTANT NOTE: As the atmosphere in an incubator is normally kept at a high humidity,
we recommend that you DO NOT leave the analyser sampling on a humidified incubator for
longer than 2 minutes. After sampling on a humidified Incubator for a long period, we
recommend that you continue to run the analyser pump in air for 30 seconds to flush out any
humidity which could have built up inside the sensor.
Pump
Pump
button to turn on the metering gas pump.
Pump
button to turn off the metering gas pump.
Pump
button.
button.