Description of the instrument

The NSG 4070 is a multi-functional device for carrying out EMC immunity tests to accompany development
and conformity testing in accordance to IEC / EN 61000-4-6, IEC / EN 61000-4-3, IEC / EN 61000-4-20, IEC / EN
61000-4-21 and several BCI standards e.g. ISO 11452-4 or MIL-STD-461G CS114.
The NSG 4070 includes signal generator, power meters, several EUT monitoring interfaces and an optional Class
A power amplifier module. The flexibility in the EMC lab is given by the wide frequency range of 4 kHz to 1 GHz,
several models of internal power amplifiers, the possibility to connect external power amplifiers and directional
couplers as well as the variety of interfaces for EUT monitoring.
The NSG 4070C1 includes extended parameters for the pulse modulation. Up to three pulse modulation settings
can be defined in order to create an envelope. NSG 4070C1 meets the requirements of ISO / DTS 7637-4 Annex
C: Test generator for pulsed sinusoidal disturbances, pulse A.
The powerful and easy to use firmware makes the NSG 4070 independent from an external PC and control soft-
ware, however it can also be remote controlled for system operation. A state-of-the-art data transfer of test and
measurement data for report generation is provided by USB stick to be plugged into the front panel.
In order to start with predefined parameter settings is recommended the optional test software icd.control. The
software offers a large standard database and predefined drives for using external measuring devices. More
complex systems including radiated tests can be controlled by using the software solution CIS (Compliance Im-
munity Software).
The NSG 4070 is supplied with remote interfaces LAN, electrical or optical RS232 and USB.
Figure 1. - Block diagram of NSG 4070 with internal power amplifier
Quick Start and Safety Guide