Error Messages
Error Message
Er EP
Er Ad
Ernn
Ersp
Erng
udrld
ourld
15
What to Do or Check
The setup parameters loaded in nonvolatile memory
have become corrupted.
Reset all parameters to factory default with the rst
parameter.
The A/D communication is not detected. Ensure the
cable is connected to the rear of the indicator.
If problem persists, internal connectors will need to be
checked. Contact Doran Technical Support for
assistance.
Motion Error, weight reading during calibration is too
unstable. This error is displayed during zero or span
calibration attempt. Error will disappear when platform
becomes stable.
Refer to the Calibration Troubleshooting section for
additional information. If this problem persists, contact
Doran Technical Support for assistance.
The span calibration is out of range. This error is
displayed after a span calibration attempt. Place the
proper weight on the platform and press TARE to
perform the span calibration again.
Refer to the Calibration Troubleshooting section for
additional information. If this problem persists, contact
Doran Technical Support for assistance.
The span calibration is in a negative range. This error
is displayed after a span calibration attempt. Place
the proper weight on the platform and press TARE to
perform the span calibration again.
Refer to the Calibration Troubleshooting section for
additional information. If this problem persists, contact
Doran Technical Support for assistance.
The scale is in underload. The load on the scale
platform is less than the Calibration zero level by
more than 20% of the scale capacity.
Add table top to scale platform. If problem persists,
recalibrate. If this problem still persists, contact Doran
Technical Support for assistance.
The scale is in overload. The load on the scale
platform exceeds the scale capacity by more than
105%.