4.    Set the HP 8502A Input Attenuation knob to 0 dB. 
 
5.    Adjust the HP 8754A trace to the top graticule line using the CH 1 REFERENCE POSITION knob. 
Turn the CH 1 REFERENCE POSITION off. 
 
6.    On the HP 8754A, turn on the CH 1 REFERENCE OFFSET. Adjust the trace, reference line, to the top CRT graticule 
using the REFERENCE OFFSET knob. The HP 8754A is now calibrated for making return loss measurements relative to 
the short. 
 
7.    If an HP 8750A Storage Normalizer is used, press the following keys on the HP 8750A: [CH 1], [INPUT], [STORE 
INPUT], and [INPUT‐MEM]. The displayed reference line will shift to the middle of the screen. Errors in the reflection 
test set will now be normalized. 
 
8.    Remove the SMA Coaxial Short and replace it with the Cokeva 11940C  probe. The displayed trace should be 16 dB 
or more below the reference line at all frequencies. If the trace is less than 16 dB below the reference line, the probe 
fails the test. 
     
 
 
5‐1.     SMA ROTARY JOINT INSTALLATION PROCEDURE   
 
 
Figure 10. SMA ADAPTER LOCATION 
 
 
 
 
 
Figure 11. RF Rotary Joint Mounted on Probe  
 
 
 
12