Operating instructions

DC current measurements
1. Set the rotary switch to 20A DC or 80A DC
2. Connect the negative output of the clamp (black) to the neg-
3. Switch the measuring instrument on and set it to 200 mV
4. With firmly closed jaw, press the zero button until the in-
5. Open the jaw and fully enclose the current-carrying conduc-
6. Read the measured voltage value and convert it to the corre-
AC current measurements
1. Set the rotary switch to 20A AC or 80A AC
2. Connect the negative output of the clamp (black) to the neg-
3. Switch the measuring instrument on and set it to 200 mV
4. Open the jaw and fully enclose the current-carrying conduc-
5. Read the measured voltage value and convert it to the corre-
2
Current Clamp CC-80
Amplicon.co.uk
Sales:
+44 (0) 1273 570 220
All manuals and user guides at all-guides.com
ative side of the input of the measuring instrument. Connect
the positive output of the clamp (red) to the positive side of
the input of the measuring instrument.
DC, 400 mV DC or 2 V DC full scale input range.
struments reads zero.
tor(s) and close the jaw again, no air gap is allowed betewen
the two halves. The output voltage is positive when the cur-
rent flows from the upside to the underside of the clamp.
sponding current value:
20 A range: 100 mV equals 1 A
80 A range: 100 mV equals 10 A.
ative side of the input of the measuring instrument. Connect
the positive output of the clamp (red) to the positive side of
the input of the measuring instrument.
AC, 400 mV AC or 2 V AC full scale input range.
tor(s) and cloze the jaw again, no air gap is allowed betewen
the two halves.
sponding current value:
20 A range: 100 mV equals 1 A
80 A range: 100 mV equals 10 A.