This is not an easy task and should not be attempted by the user; please contact Amplitude
for service.

2.1.7. PulseShaper not set

If the PulseShaper was powered down, it loses its configuration and by default blocks the
pulses. It must be set according to the "Load PulseShaper configuration" procedure. In case of
doubt whether the shaping is applied or not, reduce the regen pump to 53 % and apply the
spectral shaping again.

2.1.8. External devices

Please check that the following devices are functioning properly
-
High voltage switch not turned on: the unit is powered and the HV switch on the front
panel is ON.
-
High Voltage power supply: the failure LED is off, if on see section 1.4.
-
The embedded computers are on and responding: are the green LEDs on their front
panel is on (if not press the button and wait a minute for start-up), and check in the
embedded modules of the control software that their status is not frozen (graphs,
reading values not moving). If so try clicking "Restart" in the "Information tab, if it
doesn't work reset the computer by pressing the front button for 7 seconds and
restarting it, wait a minute for start-up.
-
The ethernet switch is ON: status LEDs of the cables are blinking while the embedded
computers are on. If the LEDs are all off power the switch off and on again, if it remains
off then contact Amplitude for service.

2.2. Hard checks

If none of the above apply, additional checks can be performed but they require opening the
amplifier and advanced knowledge is then needed; please contact Amplitude for assistance.

2.2.1. Input seed pulse in the regen amplifier

If all the checks from sections 2.1.2 (oscillator train visible on an oscilloscope), 2.1.3 ('Pockels
monitor' and 'Pulse picker monitor' signals have the duration indicated in the software) and
2.1.7 (spectral shaping applied) were performed the seeded pulses should be delivered to the
regen case unless an optical fibre from the oscillator or from the PulseShaper to the regen got
damaged. Nevertheless, it is possible to see if the pulses are sent through.
1. Stop the Front-end synchronisation signal in the SelectPulse's outputs to avoid any risk
of unwanted amplification.
2. Open the regen amplifier case and locate the output of the blue fibre from the
PulseShaper.
3. Block the beam path after the two first 45° mirrors.
4. At this point it is impossible to see the beam as the repetition rate is too low, to send
the whole oscillator train, go to the Magma 25 module / "Synchronisation" tab and
switch the "Pulse Picker" control from "Picker" to "Alignment". Now the beam should
be visible with an IR card or aa IR viewer.
5. Don't leave the pulse picker in alignment mode for too long as the PulseShaper heats
up with the incident power.
6. Don't forget to remove the beam block from the path before closing the regen case.
6
Magma 200
Quick troubleshooting guide
Version 1.00
11 avenue de Canteranne, Cité de la Photonique - Bâtiment MEROP A 33600 Pessac - France
Headquarters
/
Capital social: 317 100 €
Amplitude Laser Group
/
Tel: +33 (0)5 56 46 40 60
Headquarters
Amplitude Laser Group
Tel: +33 (0)5 56 46 40 60
/
Siret 434 637 005 - APE 6420Z
/
Fax: +33 (0)5 56 46 06 94
/ [email protected] / www.amplitude-laser.com
/
11 avenue de Canteranne, Cité de la Photonique - Bâtiment MEROPA 33600 Pessac - France
Capital social : 320 925 €
/
/
SIRET: 434 637 005 - APE 6420Z
/
Fax: +33 (0)5 56 46 06 94
/
N°TVA Intracom: FR 70 434 637 005
/
/
N°TVA Intracom: FR 70 434 637 005
/ www.amplitude-laser.com