Amprobe GROUND PROBE GP-1 Manual del usuario - Página 12

Navegue en línea o descargue pdf Manual del usuario para Equipos de prueba Amprobe GROUND PROBE GP-1. Amprobe GROUND PROBE GP-1 20 páginas.

Amprobe GROUND PROBE GP-1 Manual del usuario
Electrode Resistance:
Test C & D
8) The possible results are as follows:
• If the test rods are making good contact with the soil, all connections are correct, and the
actual value is less than 2kΩ, the measured value (RG) will be displayed in ohms (Ω).
Continue to Step 9.
• If the test rods are making good contact with the soil, all connections are correct, and
the actual value is greater than 2kΩ, "o.r." (over range) will be displayed. Recheck
connections, verify rod to soil continuity and treat if necessary, and/or use longer test rods.
9) Note the value of Rg, or press the SAVE button.The value of Rg at P 2 @67% will be
saved
to the next available memory location. Note the memory location.This is your elec-
trode or grid system resistance.
BLACK
C1
P1
Perform next steps to assure accuracy of the measurements
10)Remove P2 (RED) from the 62% rod and clip to the 52% rod.
11)Press and release the START button.
12)Note the value of Rg, or press the SAVE button.The value of Rg will be saved in the
next available memory location. Note the memory location.
Page 12
(cont'd.)
BLUE
52%
62%
RED
GREEN
P2
C2
72%
100%