DPM TT 570 A Manual de instrucciones - Página 13

Navegue en línea o descargue pdf Manual de instrucciones para Instrumentos de medida DPM TT 570 A. DPM TT 570 A 16 páginas. Micromanometer

 
Pressure:   
 
Pa 
Pa 
mbar 
0.010 
mmH
0.102 
2
inH
0.004 
2
‐3
mmHg  7.501 x 10
‐4
inHg  2.953 x 10
‐4
PSI  1.451 x 10
 
 
Volume: 
 
3
 
m
/sec 
3
m
/sec 
3
m
/hr 
3600 
l/sec 
999.97 
cfm 
2118.88 
 
Velocity:   
 
m/sec 
m/sec 
ft/min   
196.85 
 
 
Air Velocity Calculations using S.I Scales:
 
For non‐standard air conditions:
 
V  =  Velocity in m/sec
 PT  =  Pitot Tube Factor (for Ellipsoidal type 1.000) 
 B  =  Barometric pressure in mbar
 
=  Absolute temperature in °K (= t in °C + 273 where t = airstream temperature)
 
Pv  =  Velocity pressure in Pa
 
 
   
Air Velocity Calculations using Imperial Scales:
 
For non‐standard air conditions: V = 4006 x PT   30  x T x Pv
 
V  =  Velocity in ft/min
 
PT  =  Pitot Tube Factor (for Ellipsoidal type 1.000) 
 
B  =  Barometric pressure in inHg
 
T  =  Absolute temperature in °R (= t in °F + 460 where t = airstream temperature)
 
Pv  =  Velocity pressure in wg
 
CONVERSION TABLES
 
mbar
mmH
O
2
100.0
9.806
1
0.098
10.20
1
0.402
0.039
 
0.750
0.074
 
0.029
2.895 x 10
 
0.014
1.423 x 10
 
 
3
m
/hr
l/sec
0.0002
0.001
1
3.6
0.2777
1
0.5885
2.1189
 
 
ft/min
0.005
1
 
V = 1.291 x PT    1 013.25 x T x Pv
inH
mmHg
2
249.1 
133.3
2.491 
1.333
25.40 
13.60
1
0.535
1.868 
1
‐3
0.074 
0.039
‐3
0.036 
0.019
cfm 
0.0004 
1.699 
0.4719 
1
 
B
293
  B 
528
TT Series LR Manual 
Page 11
inHg
PSI
3385
6892
33.85
68.92
345.42
702.8
13.51
27.67
25.64
51.70
1
2.305
0.4338
1