2.2
COMPRESSOR INLET FILTER
The compressor inlet filter may have been removed to protect from shipping
damage. If so, install the compressor inlet filter prior to starting up the unit.
8
Oxygen is a powerful oxidizing agent. Material that burns in air and
even some material that will not burn in air are easily ignited and
burn rapidly in high concentrations of oxygen. For safety concerns,
all possible sources of ignition must be kept away from the oxygen
concentrator and preferably out of the room in which it is being
used. Observe strict cleanliness procedures when fabricating and
connecting the oxygen piping. It is imperative that oxygen systems
be properly cleaned and inspected to insure that no combustible
materials remain in the system. If you are not familiar with oxygen
cleaning procedures, refer to the Compressed Gas Association
documents G-4.1 "Cleaning Equipment for Oxygen Service" and
G-4.4 "Industrial Practices for Gaseous Oxygen Transmission and
Distribution Piping Systems."
Do not allow the free flow of oxygen from the Workhorse Oxygen
Concentrator. Ensure that the oxygen flow is measured and
controlled to rates that do not exceed rated capacity.