Operating instructions
1. Adding Water: it can be used when 3.5L water has been added
into the container, but the water level can't exceed the three
shelving plane which is under inner barrel. Water rating must be supply enough
during circulate use, in order to prevent damaging the electric heating tube due to
water shortage.
2. Heaping objects:Put these objects needed to be sterilized on sieve plate, there
should be gaps among these packed objects, to help steam to permeate and improve
working quality.
3. Air proofing: put the sanitizing cylinder into the container, and then insert the soft
tube which is on container lid into the semicircle groove in the cylinder,screw the
container lid with wing-shaped nut in container tightly, make no distance between
the lid and container.
4. Heating : connect the power source to power source; venting valve and safety
valve should be raised when heating begin to make cold air in sanitizing cylinder go
out; when temperature inside is more than 100℃(pressure 0.05MPa),and when rapid
steam jet out, restore the hats of the two valves to original.
5. Cooling:When finishing sterilizing, turn off power first, and then open the vapor
valve and the lid when the reading of pressure gauge turns back to zero after cooling.
Be cautious to not open valve to safety valve with
tube or open the lid as soon as
finishing satirizing, to avoid seething lotion overflowing which may cause accident.
- -
4