Drawell YP Series Manual de instrucciones - Página 9

Navegue en línea o descargue pdf Manual de instrucciones para Scales Drawell YP Series. Drawell YP Series 11 páginas. Industrial precision balance
También para Drawell YP Series: Manual de instrucciones (7 páginas)

Also, carry out the adjustment at a location where there are few people moving
around and there is no air flow or vibration.
1 Press "CAL"
The weight value will flash.
The calibration record is being output. When the output has finished, span
calibration will start automatically
2 Place the calibration weight on the pan.
Wait until the flashing weight value display changes to a steady 500.00.
3 Take the weight off the pan.
"0.00" will be displayed and the balance will return to the weighing mode
9