BD BACTEC MGIT 320 Manual de consulta rápida - Página 9

Navegue en línea o descargue pdf Manual de consulta rápida para Equipos de laboratorio BD BACTEC MGIT 320. BD BACTEC MGIT 320 10 páginas.

BD BACTEC MGIT 320 Manual de consulta rápida
Meaning
Code
E35
Incorrect drawer
E37
Barcode scan order incorrect
E38
No storage media in drive
E39
Storage media full, insert another
E40
Network error
LIS parameter changed:
E43
instrument must be restarted.
E45
Drawer error
E50
Internal software error
E92
AST Set Error
MA0120(11) (2022-04) • English
(If the recommended actions do not correct the situation, refer to the User's Manual or contact BD)
Tube just scanned belongs in another drawer. Open next drawer and scan tube barcode. Repeat until error
does not recur.
System expected a barcode sequence number and an accession barcode was scanned (or vice versa). Be sure
to scan the type of barcode the system expects.
Insert blank, formatted, write-enabled storage media in the storage media drive to save data to storage media.
Insert blank, formatted, write-enabled storage media in the storage media drive to complete the save data to
storage media function.
Start BD EpiCenter™ system.
Select OK soft key to implement changes or select not OK soft key to cancel changes.
A drawer is not fully closed. Close drawer.
System encountered a software error. Save data to storage media and call BD.
Main error code for any station error in an AST set. Refer to User's Manual for additional information.
Error Codes E35 – E92
Corrective Action
9