PPE must be worn at all times in our BSL-2 environment
*Lab coat, gloves, and closed toed shoes are mandatory; respirator and goggles are available*
An Instrument Configuration is located on the website in the
All samples must be in POLYPROPYLENE round bottom tubes.
What samples do I need to bring?
• All users ALWAYS need to bring single stained compensation controls
• If you have a 4 color panel you should bring: 5 compensation controls
o One tube stained with each individual antibody
o One unstained tube
o No PI or viability dye in the single color controls
Don't have enough cells for compensation?
User can purchase compensation beads from many different suppliers
• eBioscience UltraComp Beads 01-2222-42 (for panels including UV and Violet dyes)
• eBioscience OneComp Beads 01-1111-42 (for panels that do NOT include any UV or Violet dyes)
• BD CompBeads 552845 'Anti-Rat/Hamster'
• BD CompBeads 552843 'Anti-Mouse'
• ThermoFisher Amine Reactive Compensation Beads A10628 (for LIVE/DEAD Fixable stains)
lasted updated by MW on 11/10/2017
BD FACSJazz Users Guide
Satchmo
Stanford Shared FACS Facility
Instrument Overview area
*The configuration lists the lasers and filters for each machine*