AND GR Series Manual de instrucciones - Página 12

Navegue en línea o descargue pdf Manual de instrucciones para Scales AND GR Series. AND GR Series 20 páginas. Density determination kit applicable analytical balance models bm/gh/hr-i/gr/hr/ha series

5.3. Assembling the Kit: HR Series

Assembled view
Thermometer clamp
1.
Remove the weighing pan and breeze break from
the balance.
2.
Place the density pan stand on the balance.
3.
Place one small balance weight and one large
balance weight on the density pan stand.
4.
Place the beaker stand so that it does not touch
the density pan stand.
5.
Attach the thermometer clamp to the beaker
and
insert
thermometer clamp.
6.
Pour a liquid with a known density (such as
distilled water) into the beaker and place the
beaker on the beaker stand.
7.
Place the density pan on the density pan stand.
8.
Adjust the amount of the liquid so that the sample,
when placed on the lower pan (in the liquid), is about
10 mm below the surface of the liquid.
9.
When the value displayed on the balance is stable,
press the RE-ZERO key to set the displayed value
to zero.
-e
10.
When
on the balance, perform a zero calibration.
Now the balance is ready for density measurement.
Thermometer
Beaker
Beaker stand
the
thermometer
or
- - - - -
is displayed
Density pan
Density pan stand
Balance weights (1 small, 1 large)
Weighing pan
Breeze break
into
the
10