prompted to take a picture of yourself while taking the test. This may be useful
for self-accountability or verification purposes. You can disable 'camera mode'
under the mobile app's 'Settings' tab.
6. After the test is completed, the app will display the test results. You can then
add location information, leave comments about the test, or share the results
on your phone via SMS, email, or social media.
Warning Display
FLOW
The breath sample was not strong enough for a measurement. Please
take another test and provide a strong, steady, continuous blow during the
sampling process. Do not stop blowing until the beep stops and the device
makes two 'clicks'.
Bat
The battery power is low. Please replace the batteries.
CAL
If you test over 1,000 tests, the device will display the "CAL" sign for
3 seconds, and then proceed to warm up for testing. Please send the
breathalyser back for calibration.
Err1
Pressure error sensor. Please contact us at 1300 800 200.
Err2
Sensor error. Please contact us at 1300 800 200.
Err3
Out of operation temperature range - please check the ambient temperature
and operating temperature. If the problem still persists, please contact us
at 1300 800 200.
How to remove the battery cover
1. Locate the groove, which is located on the
back of your device.
2. Slide your finger into the groove, and gently
pull the battery cover off.
3. Please note that you should only use
regular, non-rechargeable, non-heavy duty
alkaline batteries in your device.