4. Maintenance and troubleshooting
4.1.
Problem
My device turns itself off.
My device is blocked, the force does
not vary anymore.
My device does not display the
measurement screen but an
information page with Offset in red.
4.2.
Reference
CNR FT 10
CNR FT 25
CNR FT 50
CNR FT 100
CNR FT 250
CNR FT 500
Main faults and solutions
Solution
Turn off and restart your instrument using the on/off key.
If the offset value is greater than 20%, the device cannot
measure. Turn off your device and remove the load
applied to the sensor. Then turn on your device.
If no load is applied to the sensor, then the sensor is
deformed. You must then return your device to us for
replacement of the sensor.
Technical specifications
Capacity
10 N
25 N
50N
100 N
250 N
500N
1. Check that the battery percentage is not at 0%.
Put the device on charge
2. The Centor First II will automatically shut off after
15 minutes if no button is pressed.
Resolution
0.002 N
0.005 N
0.01 N
0.02 N
0.05 N
0.1 N