Alert J4X Manual de instrucciones - Página 14

Navegue en línea o descargue pdf Manual de instrucciones para Equipos de prueba Alert J4X. Alert J4X 16 páginas. Breath alcohol tester

Alert J4X Manual de instrucciones

AMBIENT TEMPERATURE

Operate the ALERT J4X breath alcohol tester within the
ambient temperature range of -20 to 50 °C.
When the ambient temperature is below 15 °C, the Ambient
Temperature icon is displayed. Use the car adapter (auxiliary
power) to provide an extended temperature range to -20 °C.

TROUBLESHOOTING

The ALERT J4X breath alcohol tester requires calibration
each year to maintain accuracy. When calibration is due,
CAL is displayed.
INCIDENT / MESSAGE
No Power
No Test
CAL
E01
E02-E05, E15, E16
E06, E07
E08
E09-E25
RESOLUTION
Remove the plastic tab from the
battery compartment
Review the "Taking a Breath Test"
section. Contact ACS for assistance
Calibration is due. Contact ACS for
service
Change the batteries
Power the tester off and power back
on. If the problem persists, contact
ACS
Temperature error – operate the
tester within an ambient temperature
range of -20 to 50 °C
Temperature error – use the car
adapter. If the problem persists,
contact ACS
Contact ACS for service
ALERT J4X