Droper Field 1000 Manual de instrucciones - Página 8

Navegue en línea o descargue pdf Manual de instrucciones para Equipamiento médico Droper Field 1000. Droper Field 1000 16 páginas. Mechanical pressure infuser
También para Droper Field 1000: Manual de instrucciones (15 páginas)

2.5
Draining the infusion line and bag
Two reverse purge air drainage scenarios are possible:
1.
Draining an infusion bag outside of the Droper by exerting upwards hand pressure on the bag held
upright and perforated with the infusion line.
b.
Case where the drip chamber cannot be used upright (case 2.3 1-b described above) and
case where the line does not have a drip chamber: with both the clamp and Luer Lock cap
at the end of the line open, expel all air contained in the infusion bag until the drip chamber
is entirely filled with the infusion solution. Close the choke. Position the bag in the Droper
Field as described in point 2.4 above. Close the Droper Field, re-open the choke and drain
all residual air from the infusion line.
a.
Case where the drip chamber may be used in the
upright position (case 2.3 1-a described above): with the
clamp closed, expel all air contained in the infusion bag until
the drip chamber becomes filled with part of the liquid.
Then insert the bag into the Droper Field as described in
point 2.4 above, ensuring that the drip chamber is
positioned with the outlet pointing downwards. Close the
Droper Field, open the clamp and the Luer Lock cap located
at the end of the line and expel the residual air in the
infusion line.
8 / 16