Droper Field 1000 Manual de instrucciones - Página 11
Navegue en línea o descargue pdf Manual de instrucciones para Equipamiento médico Droper Field 1000. Droper Field 1000 15 páginas. Mechanical pressure infuser
También para Droper Field 1000: Manual de instrucciones (16 páginas)
9. OPERATING ISSUES
Stop or absence of administration due to a device malfunction
If the administration stops due to a device malfunction, characterized by a lack of bag pressure, this
incident may be confirmed by an absence of flow of the product in the drip chamber if this one is
used, or – if not used - by observance of blood reflux in the tubing when the device is located at the
patient's level or even below.
Firstly check that the cover of the Droper Field has been closed correctly. Where necessary, open and
re-close the cover.
In the event that the cover is closed correctly, open the Droper Field, unfasten the mobile plate from
its tightening clips (see point «Cleaning » in point « 10. CLEANING / DISINFECTION »), check the
following points and implement the relevant actions :
Check that there are no objects blocking the mechanism. If this is the case, remove the object
Reinstall the mobile plate onto the mechanism by replacing the clips using vertical force.
Re-pressurise the system.
Check that one or several springs have not stopped working
Remove the bag of fluid and continue administration by gravity or manual pressure on
the bag, or change device. Send the Droper Field to the maintenance department for repair.
Check whether the roller clamp or flow regulator is not set to the OFF position
Reposition the roller clamp or flow regulator to the desired position.
Check whether the infusion / transfusion line is not being pinched by excessive bending of the
latter
Remove this bend.
Check whether the infusion / transfusion line is not being clamped by an object crushing the
line
Remove this object.
Check whether the infusion / transfusion line has not been clamped between the cover and
the plate when arming the Droper Field
Open the Droper Field and move the infusion / transfusion line.
Stop, insufficient or absence of administration due to a catheter obstruction
After checking the correct operation of the device, check whether the administration has stopped, is
insufficient or is absent by observing the slowing or absence of flow in the drip chamber, or - if the
drip chamber is not used - by the absence of blood reflux in the tubing if the Droper Field is disarmed
and installed at the patient's level or even below. The issue is then generated by a catheter
obstruction.
Reinsert another catheter.
Instructions for use - Version 2
11/15