Troubleshooting
Problem
No response when > is clicked
in Visulux software.
Beam plot and surface picture
appear truncated at the top of
the profile.
No response when > is
clicked in Visulux software.
Appendix A: Changes to Manual
Revision Date
Revision #
4-8-10
C
1/10/17
D
1 2
Possible Cause
Camera is not responding.
Camera is saturated and cannot
respond properly.
Camera is not selected as input
device.
Summary
Updated assembly drawing
General updates
Solutions
Check the serial number of camera and
open the Visulux software folder to check
for the presence of the proper key. There
may be more than one key present in the
software folder, and the key ID may be a
bit hard to decipher but check the camera
serial number and it should be close to
one of the keys present. The master CD of
the Visulux software will contain the
proper key file.
Reduce laser power or reduce exposure
time.
On the toolbar, go to View (if Input
Control is not checked, check the box to
bring up this window, which will be titled
Camera). Under Select a video board, the
camera should appear here and be
highlighted. If it is not, highlight this.