Ankom SDF Procedimiento
Navegue en línea o descargue pdf Procedimiento para Instrumentos de medida Ankom SDF. Ankom SDF 2 páginas. Ash determination procedure
Ash Determination
IDF
To determine the ash content within the IDF residue, follow the procedure below.
1.
Starting at the bottom of the filter part of the bag, fold the bag upward, leaving at least 10 mm at the top of the
clear polypropylene that is not folded. Then fold the bag in thirds from side to side.
10 mm of clear
polypropylene
With your Heat Sealer set to "6" (setting may vary depending on heat sealer and power source), press the Heat
2.
Sealer arm down for 3 to 4 seconds to seal the clear polypropylene part of the bag just above the white filter
material.
3.
Label the folded, sealed filter bag and place it in a labeled ashing crucible.
Repeat steps 1 – 3 for all IDF bags to be ashed.
4.
5.
Place covered ashing crucibles with filter bags in a furnace at 600°C for 3 hours.
6.
After the 3 hour furnace time, place the ashed crucibles in desiccator(s) and allow them to cool.
7.
Remove cooled crucibles from the desiccator(s) and record their weights.
Rev 9/20/13
Ash Determination Procedure – IDF/SDF/TDF
ANKOM Technology recommends that you use ceramic crucibles with fitted
covers to avoid cross contamination or loss of ash. Fold the filter bags so that they
easily fit into the crucibles.