Introduction
The IDOS Universal Pressure Modules (UPM) use
Intelligent Digital Output Sensor (IDOS) technology to
measure the applied pressure and supply the data to an
IDOS instrument.
The IDOS technology gives instant plug and play
functionality with all instruments that have the IDOS facility.
The power for the UPM comes from the IDOS instrument.
For the full specification and user manual, refer to Druck
website:
WARNING
Some liquid and gas mixtures
are dangerous. This includes mixtures that
occur because of contamination. Make sure
that the UPM is safe to use with the necessary
media.
It is dangerous to ignore the specified limits
(refer to data sheet) for the UPM or to use the
UPM when it is not in its normal condition. Use
the applicable protection and obey all safety
precautions.
To prevent a dangerous release of pressure,
isolate and bleed the system before
disconnecting a pressure connection. A
dangerous release of pressure can cause
injury.
Do not apply pressure greater than the
maximum safe working pressure.
Do not apply pressure greater than that
specified in Table 5 to the REFERENCE
pressure port.
Safety
The UPM has been designed to be safe when operated
using the procedures detailed in this manual. Do not use
this equipment for any other purpose than that stated, the
protection provided by the equipment may be impaired.
Before installing and using the UPM, read and understand
all the related data. This includes: all local safety
procedures and installation standards, and this document.
Repair
Do not do repairs to this equipment. Return the equipment
to the manufacturer or an approved service agent.
Druck.com
Symbols
Symbol
Description
This equipment meets the requirements of all
relevant European safety directives. The equipment
carries the CE mark.
This symbol, on the equipment, indicates a warning
and that the user should refer to the user manual.
Ce symbole, sur l'appareil, est un avertissement qui
indique que l'utilisateur doit consulter le manuel
d'utilisation.
Do not dispose of this product as household waste.
Use an approved organization that collects and/or
recycles waste electrical and electronic equipment.
For more information, contact one of these:
- Our customer service department: Druck.com
- Your local government office.

Installation

WARNING
Only fluids that are compatible
with Stainless Steel and Hastelloy shall be
used with the pressure sensors. This is to
ensure the integrity of the pressure sensor
and avoid fluid leakage.
CAUTION
To prevent damage, do not apply
torque to the body of the UPM. If available, use
the flat faces on the pressure connector to
hold the UPM in position.
For an overview of the equipment connections, refer to
Figure A1 and the explanation below:
1. Pressure port.
2. Cable strap.
3. UPM cable. For storage, wind the cable in the direction
shown.
4. Reference pressure port on a gage (g) and differential
(d) sensor. PTFE vent on sealed gage (sg) or absolute
(a) sensor. Refer to Table 3.
5. Communications port connector for an IDOS
instrument. The connector includes a thread to lock the
connector in position.
Pressure Connection
Before starting, refer to the MWP (maximum working
pressure) rating specified on the product label.
Pressure port connections for G1/8 fittings are shown in
Figure C1.
1. UPM G1/8 pressure connector.
2. Bonded Seal, e.g. Dowty 400-003-4490-41 or
equivalent.
3. ISO 228/1 G1/8 pressure connector.
4. NPT pressure connector.
5. NPT female to G1/8 male adapter IO-ADAPT-1/4NPT
or IO-ADAPT-1/8NPT.
For pressures less than 100 bar (1450 psi) when using
G1/8 fittings, see alternative sealing method in Figure C2
and the key below:
1. UPM G1/8 pressure connector.
2. Bonded seal.
3. ISO228/1 G1/8 pressure connector or adaptor.
Pressure port connections for NPT fittings are shown in
Figure C3.
1. UPM 1/8 NPT pressure connector.