Drucker Diagnostics Horizon 12 Manual de servicio - Página 8

Navegue en línea o descargue pdf Manual de servicio para Equipos de laboratorio Drucker Diagnostics Horizon 12. Drucker Diagnostics Horizon 12 12 páginas.

Drucker Diagnostics Horizon 12 Manual de servicio
Fig 3
Insert the Elastomer bushings into the motor mount plate as shown (fig 4).
Push the grommets onto the bushings as shown (fig 5).
Fig 6
Place one Aluminum spacer 30-000112 onto each stud as shown (fig 6).
Install the motor mount assembly onto the guard bowl studs so that the motor
wire tiedown is positioned opposite of the guard bowl splice (fig 7).
Install one fender washer onto each stud (fig 8).
Finish the assembly with one Nylok nut on each stud, and tighten to 13 in-lbs
DRUCKER DIAGNOSTICS
HORIZON 12 SERVICE MANUAL
Fig 4
Tiedown
SPLICE
Fig 7
SM015
REV A
Fig 5
Fig 8
8