Drucker Diagnostics Horizon 6 Flex Manual del usuario - Página 6

Navegue en línea o descargue pdf Manual del usuario para Equipos de laboratorio Drucker Diagnostics Horizon 6 Flex. Drucker Diagnostics Horizon 6 Flex 14 páginas.
También para Drucker Diagnostics Horizon 6 Flex: Manual de servicio (10 páginas)

Drucker Diagnostics Horizon 6 Flex Manual del usuario
I
S
NITIAL
ETUP
o Unpack and verify that all the following are included:
o Setup the centrifuge on flat and level surface. A bench top clearance height of 21" (54 cm) is required to open
the lid.
o The centrifuge should have 6" (15 cm) of clear space around the centrifuge. Proper ventilation is necessary to
prevent the overheating of samples as well as premature failure of the centrifuge. Choose an area which allows
unencumbered air flow, and where the temperature remains between 16°C and 32°C.
o No hazardous material shall be permitted in the clearance envelope during operation.
o The operator time within the envelope shall be limited to the time necessary for loading, unloading, and
centrifuge operation only.
o Plug the line cord into the centrifuge.
o Plug the line cord into an approved electrical outlet.
o Turn on the power switch in the back of the centrifuge.
B
E SURE THE ELECTRICAL OUTLET IS ALWAYS ACCESSIBLE AS THE LINE CORD IS THE MEANS OF EMERGENCY
D
!
ISCONNECTION
O
PERATION
o Place the tubes into the tube holders. Be sure to follow the rules for balanced loads as listed in the next section.
o Close the lid and turn the lid knob clockwise to its complete stop position.
o The digital screen shows the currently selected cycle. To select another cycle, press the UP or DOWN button in
succession until the desired cycle is selected.
o Pushing the START button on the control panel will start the spin cycle.
o When the cycle is completed, the rotor will slow to a complete stop and the lid light will flash.
o The unlocking mechanism will engage for 60 seconds allowing entry into the rotor chamber. To unlock after more
than 60 seconds have elapsed, press the UNLOCK button. The lid will unlock for another 15 seconds.
o Turn the lid knob counterclockwise and open the lid. The lid light will turn off.
Drucker Diagnostics – Customer Service: +1-814-692-7661 –
Centrifuge
Power cord
Tube holders
Tube Inserts
Tube Holder Caps
Quick Start Guide
| Page 4